26
the pump pressure is higher than 300 mmHg (
appears on the display)
pumping up takes longer than 160 seconds (
appears
on the display)
there is a system or device error (
, , or
appears on the display)
the batteries are almost empty .
In such cases, repeat the measurement. Ensure you do not
move or speak.
If necessary, re-insert or replace the batteries.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section “Technical speci-
fications”, the technical alarm will appear on the display
indicating either “
” or “Lo. In such cases, you should seek
medical assistance and check the accuracy of your proce-
dure.
The limit values for the technical alarm are factory set and
cannot be adjusted or deactivated. These alarm limit val-
ues are accorded second priority under the standard IEC
60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear auto-
matically after about 8 seconds.
9. Cleaning and storing the instrument
Clean your device and cuff carefully only with a slightly
moistened cloth.
Do not use detergents or solvents.
On no account must you immerse the computer in water,
otherwise liquid can enter it and cause demage.
When storing the device, make sure that no heavy objects
are placed on top of it. Remove the batteries. The cuff
tube should not have any sharp kinks.
10. Specifications
Model no. BM 58
Measurement
method
Oscillometric, non-invasive blood pres-
sure measurement on the upper arm
Measurement
range
Cuff pressure 0 300 mmHg,
systolic 60 – 260 mmHg,
diastolic 40 –199 mmHg,
Pulse 40 –180 beats/minute
Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
inaccuracy
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory 2 x 60 memory spaces