Beurer GL32 mg/dL
29
7 STORAGE AND MAINTENANCE
Storing your device
After use, store the measuring device in the supplied case.
Note
• Keep these instructions for use in a safe place.
• If you do not use the device for a prolonged period of time, remove the
batteries.
Maintenance
Clean the surface of the device using a damp cloth (water or mild detergent). Dry the
device using a lint-free cloth.
Note
The measuring device is made from precision components. The accuracy of
the measured values and the service life of the device depend on careful
handling:
• Protect the device against impact and do not drop it.
• Protect the device against damaging influences such as moisture, dirt, dust,
blood, control solution or water, extremes of temperature and direct sunlight.
• Do not store close to strong magnetic fields, wireless systems or mobile
telephones.
8 TROUBLESHOOTING
Problem: Unexpected message on the display
No. Cause Action
E-b Flat batteries Replace all batteries.
E-F Test strip was removed from the
device during measurement
Repeat the blood sugar measurement
using a new test strip
E-t Temperature of the measuring
environment, measuring device
or test strip was outside the
permissible range
Repeat the test with a new test strip once
the measuring environment, measuring
device, and test strip have reached room
temperature (+20°C to +25°C).
A used or contaminated test
strip was inserted
Insert an unused, clean test strip and repeat
the measurement
E-U
Blood was applied to the test
strip too early
Repeat the test using a new test strip. Do
not apply the blood until the blood droplet
symbol 6 flashes in the display.