Beurer GL44 mg/dL 25
Warning
The result range printed on the test strip box applies only to the control solution. This is not a rec-
ommended value for your blood glucose level.
If measured values are outside the specified range, check the following possible causes:
Cause Action
The first drop of control solution was not dis-
posed of.
The tip of the bottle was not cleaned correctly.
The bottle was not shaken well enough.
Rectify the cause and repeat the test.
Control solution and/or test strips have passed
their expiry date or are contaminated.
Repeat the test using a new bottle of control
solution and/or new test strip.
The control solution, test strips or blood glucose
monitor are too warm or too cold.
Bring the control solution, test strips and blood
glucose monitor to room temperature (+20 °C to
+26 °C) and repeat the test.
The test strips and control solution were kept at
temperatures and in humidity outside the speci-
fied range.
Repeat the test using new/correctly stored test
strips and control solution.
Damaged test strips.
For example
Test strips that were exposed to fresh air for
too long.
The test strip box was not closed completely.
Repeat the test using a new test strip and/or a
new box of correctly stored test strips.
Old test strips. Open a new box of test strips. Repeat the test.
There is a problem with the blood glucose moni-
tor.
Contact our customer service.
Functional test was incorrectly performed. Repeat the test and follow the instructions.
Warning
Do not use the system to measure your blood glucose level if you are repeatedly provided with
measured values outside the specified range when using control solution. Contact our customer
service.