Recommended separation distances between
portable and mobile RF communications equipment and the TL90
The TL90 is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of the Medical Daylight can help prevent electromagnetic interference by
maintainingaminimumdistancebetweenportableandmobileRFcommunicationsequipment(transmitters)
and the TL90 as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of
transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d = 1.2 x P
1/2
80 MHz to 800 MHz
d = 1.2 x P
1/2
800 MHz to 2,5 GHz
d = 2.3 x P
1/2
0,01 0.12 0.12 0.23
0,1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
inmeters(m)canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter,wherePis
themaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufacturer.
NOTE 1At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorpti-
on and reflection from structures, objects and people.