Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The BM 70 is intended for use in the electromagnetic environment speci ed below. The customer or the user of the BM 70 should assure that it is
used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If
oors are covered with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
± 2 kV for power supply lines
N/A
Mains power quality should be that of a typical com-
mercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV di erential mode
N/A
Mains power quality should be that of a typical com-
mercial or hospital environment.
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
Mains power quality should be that of a typical com-
mercial or hospital environment. If the user of the BM 70
requires continued operation during power mains inter-
ruptions, it is recommended that the BM 70 be powered
from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz)
magnetic fi eld IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fi elds should be at levels
characteristic of a typical location in a typical commer-
cial or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.