Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration – electromagnetic immunity
The FT 90 is intended for use in the electromagnetic environment specied below.
The customer or the user of the FT 90 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment –
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or
ceramic tile. If oors are covered with
synthetic material, the relative humi-
dity should be at least 30 %.
Electrical fast transient/
burst
IEC 61000-4-4
2 kV for power supply lines
1 kV for input/output lines
Not applicable Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
1 kV line(s) to line(s)
2 kV line(s) to earth
Not applicable Mains power quality should be that
of a typical commercial or hospital
environment.
interruptions and voltage
variations on power sup-
ply input lines
IEC 61000-4-11
< 5 % UT (> 95 % dip in UT)
for 0.5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
< 5 % UT (> 95 % dip in UT)
for 5 sec
Not applicable Mains power quality should be that
of a typical commercial or hospi-
tal environment. If the user of the
FT 90 requires continued operation
during power mains interruptions, it is
recommended that the FT 90 be pow-
ered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz) magnetic eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.