Beurer GL 50 evo (mgdl) User Manual Page 19

  • Download
  • Add to my manuals
  • Print
  • Page
    / 40
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 18
Beurer GL50 evo mg/dL 19
5.2 Preparing the lancing device for taking a sample
1
Remove cap
Securely hold the monitor with one hand on the lancing device cover.
With your other hand, remove the cap from the lancing device.
2
Insert lancet needle
Insert a sterile lancet needle into the lancing device.
Push firmly on the lancet until it audibly engages and it cannot be
pushed further into the holder.
3
Remove the protective lancet disc
Remove the protective lancet disc by turning clockwise. Retain the
protective disc for the safe disposal of the used lancet needle after
taking a blood sample.
4
Select cap and attach.
You need to use a dierent cap depending on the area from which you
take the sample:
- Fingertip: Cap (black)
- Other body parts: AST cap (transparent)
Warning
Do not use the AST cap to take blood samples from the finger.
Place the selected cap on the lancing device. Make sure that the curved
part of the cap fits on the curved part of the lancing device. Press firmly
on the cap until it audibly engages.
Page view 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 39 40

Comments to this Manuals

No comments