32 Beurer GL50 evo mg/dL
Note
The blood glucose monitor is made of precision components. The accuracy of the measurements
and service life of the device depend on its careful handling:
Protect the device from impacts and do not drop it.
Protect the device from damaging factors such as moisture, dirt, dust, blood, control solution or
water, marked temperature fluctuations, direct sunlight and extreme cold.
Do not use the device in the vicinity of strong electromagnetic fields, radio systems or mobile
telephones.
8 What if there are problems?
Display messages on batteries and blood glucose measurement
No. Cause Solution
Empty
battery
Batteries discharged.
Charge the blood glucose monitor via the USB
interface.
High
temperature
Very high glucose level
(> 630mg/dL)
Temperature of the measuring environment,
blood glucose monitor or test strip above the
permitted range.
Repeat the test using a new test strip as
soon as the measuring environment, blood
glucose monitor and test strips have reached
room temperature (+20°C to +26°C).
Low
temperature
Very low glucose level
(< 20mg/dL)
Low temperature Temperature of the measur-
ing environment, blood glucose monitor or test
strip below the permitted range.
Repeat the test using a new test strip as soon
as the measuring environment, blood gluco-
se monitor and test strips have reached room
temperature (+20°C to +26°C).
Strip
error
Used or contaminated test strip
inserted.
Insert unused test strip that has not expired.
Repeat the blood glucose measurement.
Error 1 System error Remove batteries, reinsert batteries.
Contact our customer service if the problem
persists.
Error 2 Insucient amount of blood on
the test strip.
Repeat the measurement using a new test
strip.