If your smartphone has a protective cover, remove this to ensu-
re that there is no interference during the transfer. Begin the
data transfer in the Beurer “HealthManager” app.
9. Cleaning and storing the unit
Clean the unit and cuff carefully using a slightly damp cloth
only.
Do not use any cleaning agents or solvents.
Under no circumstances should you hold the unit under wa-
ter, as this can cause liquid to enter and damage the unit.
If you store the unit, no heavy objects should be placed on
top of it. The cuff line should not be bent sharply.
1
0. Error messages/trouble-shooting
In case of faults, the _ message appears in the display.
Error messages may appear if:
systolic or diastolic pressure could not be measured ( or
appears on the display)
systolic or diastolic pressure was outside the measurement
range ( or Lo appears on the display)
the cuff is fastened too tightly or loosely ( or appears
on the display)
the pump pressure is higher than 300 mmHg ( appears
on the display)
pumping up takes longer than 160 seconds ( appears on
the display)
there is a system or device error ( , , or ap-
pears on the display)
the data could not be sent to the PC ( appears in the
display).
the data could not be sent via Bluetooth
®
(
).
In such cases, repeat the measurement and/or the data trans-
fer. Ensure that the cuff line is correctly attached and that you
do not move or speak during the measurement.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section “Technical specifica-
tions”, the technical alarm will appear on the display indicating
either “
” or “ Lo . In such cases, you should seek medical as-
sistance and check the accuracy of your procedure.
The limit values for the technical alarm are factory set and can-
not be adjusted or deactivated. These alarm limit values are
accorded second priority under the standard IEC 60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear automati-
cally after about 8 seconds.
11. Battery
The BM 85 is equipped with a lithium ion battery (3.7V/400
mAh). If the symbol flashes, you must charge the battery
for a minimum of two hours using the enclosed cable. Rechar-
ge the battery to between 50 and 75 % of its capacity at least
twice a year to achieve a maximum battery service life.