Beurer GL44 mg/dL 33
Problem: the test does not start after inserting the test strip into the device and applying blood
Cause Solution
Insucient amount of blood or test strip not
filled correctly.
Repeat test using a new test strip and a larger
drop of blood. Please note the correct filling of
the test strip (see page 20).
Faulty test strip. Repeat the test using a new test strip.
Blood was applied while the device was switched
o.
Repeat the test using a new test strip and only
apply blood after and flash.
The basic settings of the device were changed
and the changes were not completed (see “Mak-
ing and changing basic settings” on page 14).
Remove the test strip and press the “On/O”
button until “ ” is displayed.
Repeat test.
Device faulty. Contact customer service.
9 technical specifications
Dimensions (W x H x D) 52 x 95 x 16 mm
Weight 44 g incl. batteries
Power supply 2 x 3 V CR 2032 button cell batteries
Battery life More than 1000 measurements
Measured value memory 480 measured values with date/time
Data retained upon changing batteries
Average values for 7, 14, 30, 90 days
Automatic switch-o Two minutes after last actuation
Storage/
transport temperature
Temperature: +2 °C – +30 °C
Relative humidity: < 90 %
Operating ranges Temperature: +10 °C – +40 °C
Relative humidity: < 90 % non-condensing
Measuring range, glucose Glucose: 20 – 630 mg/dL
Blood sample capillary whole blood
Amount of blood 0.6 microlitres
Blood glucose measuring duration Approx. 5 seconds
Calibration Plasma
Test procedure Amperometric bio sensor
Using the device Suitable for self-testing
System function test Upon every switch-on