Beurer GL40 mg/dL 27
Test strip inserted incorrectly or not completely.
Insert the test strip, contacts first, into the slot
on the unit. Ensure that the front side of the test
strip is facing you (see "Test strips", page 12).
Unit is defective Contact customer service.
Problem: After inserting the test strip into the unit and applying the blood, the test does not start.
Cause Solution
Insucient quantity of blood
Repeat the test using a new test strip and larger
blood drop.
Test strip is defective Repeat the test using a new test strip.
Blood was applied while the device was switched
o.
Repeat the test using a new test strip and only
apply blood after and flash.
Theunit'sbasicsettingswerechangedandthe
change was not completed (see "Basic set-
tings", page 14).
Press the "On/O" button until “ ” is dis-
played.
Repeat test.
Unit is defective Contact customer service.
9 technicAl specificAtions
Dimensions (W x H x D) 47 x 85 x 14 mm
Weight 43 g incl. batteries
Power supply 2 x 3 V CR 2032 button cell batteries
Battery life More than 1000 measurements
Measured value memory 480 measured values with date/time
Data retained upon changing batteries
Average values for 7, 14, 30, 90 days
Automatic switch-o Two minutes after last actuation
Storage/
transport temperature
Temperature:+2°C–+30°C
Relative humidity: < 90 %
Operating ranges Temperature:+10°C–+30°C
Relative humidity: < 90 % non-condensing
Measuring range, glucose Glucose: 20 - 630 mg/dL
Blood sample capillary whole blood
Amount of blood 0.6 microlitres
Blood glucose measuring duration Approx. 5 seconds