Electromagnetic Compatibility Information
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emissions
The BM 75 is intended for use in the electromagnetic environment specied below.
The customer or the user of the BM 75 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1
The BM 75 uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The BM 75 is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies buil-
dings used for domestic purposes.
Harmonic emissions IEC 61000-3-2 Class A
Voltage uctuations/icker emissions
IEC 61000-3-3
Complies