Table 3
Guidance and manufacturer´s declaration – electromagnetic immunity
The BC40 Wrist Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of
the BC40 Wrist Digital Blood Pressure Monitor should assure that it is used in such an environment.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
rms
150 kHz to
80 MHz
3 V/m 80 MHz to
2,5 GHz
Not applicable
3 V/ m
Portable and mobile RF communications equipment should be used no closer to any part
of the BC40 Wrist Digital Blood Pressure Monitor, including cables, than the recommen
-
ded separation distance calculated from the equation applicable to the frequency of the
t
ra
nsmitter.
Recommended separation distance:
80 MHz to 800 MHz
800 MHz to 2,5 GHz
where p is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in metres
(m).b
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,a should be less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is aected by absorption and reflection from structures,
objects and people.
P
V
d ]
5.3
[
1
=
P
V
d
]
5.3
[
1
=
P
E
d ]
7
[
1
=