23
Do not use detergents or solvents.
On no account must you immerse the computer in water,
otherwise liquid can enter it and cause demage.
When storing the device, make sure that no heavy objects
are placed on top of it. Remove the batteries. The cuff tube
should not have any sharp kinks.
8. Rectifying faults
E
rror messages can occur when
you move or talk during the measurement ( , ),
the cuff tube is not properly inserted ( ),
when changing the batteries, the start/stop button was
pressed,
inflation takes longer than 20 seconds ( ),
the inflation pressure is higher than 300 mmHg ( ).
In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do
not move or talk. Re-insert the batteries if necessary, or else
replace them.
9. Specifications
Model no. BM 60
Measurement
method
Oscillometric, non-invasive blood pres-
sure measurement on the upper arm
Measurement range Cuff pressure 0 300 mmHg,
systolic 50 – 250 mmHg,
diastolic 40 –180 mmHg,
Pulse 40 –160 beats/minute
Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
inaccuracy
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory 2 x 60 memory spaces
Dimensions L 101 mm x W 130 mm x H 54 mm
Weight Approx. 387 g (without batteries)
Cuff size 22 to 32 cm
Permissible operating
conditions
+10 °C to + 40 °C, ≤ 85 % relative
air humidity (non-condensing),
700 –1060 hPa ambient pressure
Permissible storage
and transport con-
ditions
- 5 °C to + 50 °C, ≤ 85 % relative air
humidity
Power supply
4 x 1,5 V
AA batteries
Battery life For approx. 300 measurements,
depending on the blood pressure level
and/or pump pressure
Accessories Instruction for use, 4 x 1.5 V AA batter-
ies, storage pouch
Classification Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF (cuff), CLASS II (when using a
mains part)
Maximum surface
temperature
Device: 44,1 °C
Mains part: 47,2 °C